10 Configuration
44
Minimum alarm (On-signal when value drops below lower limit)
Maximum alarm(On-signal when value exceeds upper limit)
Alarm window (On-signal within a configurable value range
)
Invert alarm window (On-signal outside a configurable value range)
1
0
Binary
value
Input signal
Limit value
Hysteresis
1
0
Binary
value
Input signal
Limit value
Hysteresis
1
0
Binary
value
Input signal
Limit value
Window range
Hysteresis
Hysteresis
1
0
Binary
value
Input signal
Hysteresis
Hysteresis
Limit value
Window range