12 Calibrating a pH measuring chain
60
4
Entry of the pH-value of the buffer solution
•
without buffer recognition:
Check whether the "pH buffer 1" matches the pH-value of the
buffer solution used. If a buffer set table was not specified, the
"buffer 1 pH" value is taken from the calibration default settings.
It is still possible to change this manually here.
•
with buffer recognition:
A prerequisite here is that a buffer set table has been selected
in the buffer set table and the pH-value of the buffer solution
used is contained in this buffer set table. If these prerequisites
are satisfied, the pH value of the buffer solution is determined
automatically during the calibration.
5
Wait until the measured value displayed stabilizes and then con-
firm the result of the measurement by pressing "OK"
6
A Protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
7
Press "Yes" to accept the Calibration values determined and enter
the Calibration in the Calibration logbook.
Press "No" to discard the results.
Step
Action
tap button to en-
ter pH-value of
the buffer
field
Summary of Contents for AQUIS touch S
Page 8: ...Inhalt ...
Page 20: ...10 Configuration 20 Chapter 10 12 Calibration timer page 46 ...
Page 68: ...13 Calibrating redox sensors 68 ...
Page 78: ...15 Calibrating CR conductivity sensors 78 ...
Page 88: ...16 Calibrating Ci conductivity sensors 88 ...
Page 96: ...17 Calibrating universal inputs 96 ...
Page 106: ...18 Calibrating O DO sensors 106 ...
Page 146: ...21 Annex 146 ...
Page 147: ......
Page 148: ......
Page 149: ......