10 Configuration
10
10.3
Basic settings
Open:
Device Menu > Configuration > Basic Settings
Configuration point Selection/
setting option
Explanation
Device name
up to 20 text characters
Device ID, e.g. for identification of exported mea-
surement data in the JUMO PCA 3000 evalua-
tion software
National language
German
English
Setting the operating language
Additional languages can be installed on the de-
vice using the setup program.
⇨
Operating manual B 202581.0
Language select after
power on
Yes
No
Specifies whether a prompt for selecting the op-
erating language should appear when the device
is switched on
Mains frequency
50 Hz
60 Hz
Supply frequency of the electric power grid serv-
ing the mounting site
The supply frequency must be specified to sup-
press EMC interference caused by the mains
voltage. Setting the correct supply frequency is
thus also necessary when supplying the device
with direct voltage.
Device temp.
Degrees Celsius
Degrees Fahrenheit
Setting the default on the temperature unit for all
temperature values in the device
Interface temperature Degrees Celsius
Degrees Fahrenheit
Setting the default on the temperature unit for all
temperature values communicated via interfaces
Memory alarm limit
0 to 100 %
If the available memory display reaches this val-
ue, the memory alarm is triggered.
Summary of Contents for AQUIS touch S
Page 8: ...Inhalt ...
Page 20: ...10 Configuration 20 Chapter 10 12 Calibration timer page 46 ...
Page 68: ...13 Calibrating redox sensors 68 ...
Page 78: ...15 Calibrating CR conductivity sensors 78 ...
Page 88: ...16 Calibrating Ci conductivity sensors 88 ...
Page 96: ...17 Calibrating universal inputs 96 ...
Page 106: ...18 Calibrating O DO sensors 106 ...
Page 146: ...21 Annex 146 ...
Page 147: ......
Page 148: ......
Page 149: ......