109
19 Calibrating turbidity sensors
19.3
Calibration routines for turbidity sensors
19.3.1
Two-point calibration
NOTE!
Turbidity sensors can be configured with automatic selection of the measuring
range. Accordingly, calibrations must be performed for all "accessible measur-
ing ranges".
NOTE!
You must be logged in with corresponding user rights to perform calibrations.
⇨
Chapter 8.2.1 "Log-on/Log-out", page 91
NOTE!
Digital sensors need to be linked before they can be calibrated.
⇨
Chapter 8.2.7 "Digital sensors", page 97
Step
Action
1
Start the two-point calibration.
Device Menu > Calibration > Digital Sensor 1 to 6 >
Two-Point Calibration
2
Enter one of the measuring ranges 1 to 4.. Confirm the entry by
pressing "OK"
3
Clean the sensor and immerse it in the distilled water.
4
The device displays the current measured value of the distilled wa-
ter. The user does not need to make any entries here.
5
Wait until the measured value displayed stabilizes and then con-
firm the result of the measurement by pressing "OK"
6
Immerse the sensor in a formazine reference solution. Its turbidity
value should be approximately at the middle of the selected mea-
suring range.
Summary of Contents for AQUIS touch S
Page 8: ...Inhalt ...
Page 20: ...10 Configuration 20 Chapter 10 12 Calibration timer page 46 ...
Page 68: ...13 Calibrating redox sensors 68 ...
Page 78: ...15 Calibrating CR conductivity sensors 78 ...
Page 88: ...16 Calibrating Ci conductivity sensors 88 ...
Page 96: ...17 Calibrating universal inputs 96 ...
Page 106: ...18 Calibrating O DO sensors 106 ...
Page 146: ...21 Annex 146 ...
Page 147: ......
Page 148: ......
Page 149: ......