10 Configuration
26
CR/Ci measuring range configuration
Open:
Device Menu > Configuration > Analog Inputs >
Analysis Input 1 to 4 > Measuring Range 1 to 4
Configuration point Selection/
setting option
Explanation
TDS factor
0.01 to 2.00
only for conductive conductivity with
TDS compensation:
Conversion factor from measured conductivity to
display unit (see configuration item "Unit" in this
table)
Compensation
for CR/Ci:
Off,
TC linear,
natural water,
natural water with
expanded temperature
range
Type of temperature compensation for conduc-
tivity measurement
only for CR:
TDS,
ASTM neutral,
ASTM acid,
ASTM alkaline
only for Ci:
TC curve,
NaOH 0 to 12 %,
NaOH 25 to 50 %,
HNO3 0 to 25 %,
HNO3 36 to 82 %,
H2SO4 0 to 28 %,
H2SO4 36 to 85 %,
H2SO4 92 to 99 %,
HCL 0 to 18 %,
HCL 22 to 44 %
Unit for
calculation
for CR/Ci:
µS/cm
mS/cm
Unit in which the conductivity is displayed
only for CR:
k
Ω
/cm
M
Ω
/cm
Unit
up to 5 text characters
only for conductive conductivity measure-
ment with
TDS compensation:
Unit for the process variable to be displayed for
TDS measurements or when using customer-
specific linearization (e.g. ppm or mg/l)
Summary of Contents for AQUIS touch S
Page 8: ...Inhalt ...
Page 20: ...10 Configuration 20 Chapter 10 12 Calibration timer page 46 ...
Page 68: ...13 Calibrating redox sensors 68 ...
Page 78: ...15 Calibrating CR conductivity sensors 78 ...
Page 88: ...16 Calibrating Ci conductivity sensors 88 ...
Page 96: ...17 Calibrating universal inputs 96 ...
Page 106: ...18 Calibrating O DO sensors 106 ...
Page 146: ...21 Annex 146 ...
Page 147: ......
Page 148: ......
Page 149: ......