20 Technical data
118
20.3
Analog outputs of base unit and optional boards
20.4
Digital inputs base unit
20.5
Digital inputs optional boards
20.6
Digital outputs, power supply unit board
Signal type
Signal range
Admissible
load resistance
Accuracy
Ambient tem-
perature in-
fluence
Voltage signal
0 to 10 V
> 500
Ω
≤
0.25 %
≤
100 ppm / K
Current signal
0/4 to 20 mA
< 450
Ω
≤
0.25 %
≤
100 ppm / K
Description
Input frequency
ranges
Min. pulse dura-
tion
Signal type
Switching thresh-
olds
a
a
This specification is relevant only if the Ext. voltage supply is selected under the “Contact” point in the
Configuration. Sensors and transmitters should be supplied from voltage supply outputs on the JUMO
AQUIS touch S. An externally supplied voltage signal must not have a voltage over 28 V.
On
Off
On
Off
IN 1
b
b
All digital inputs IN 1 to 3 are suitable for connecting proximity switches. Recommended
types are: Wachendorff P2C2B1208NO3A2 and Balluff BES M12EG-PSC80F-BP03.
≤
1 Hz
300 ms
300 ms
Configurable as:
potential-free con-
tact,
or
ext. voltage supply
(maximal 28 V)
> 8 V
> 1.8 mA
< 5 V
< 1.2 mA
IN 2 to 3
b,c
Switching sig-
nal
c
Binary inputs IN 2 and IN 3 can be used for impeller flow-through sensors (water meters) or
magnetic-inductive flow-through meters (square-wave signal), for example. The input frequency depends
on the configured measurement principle in the flow function.
≤
1 Hz
30
μ
s
30
μ
s
IN 2 to 3
b,c
Flow
3 to 300 Hz
300 Hz
to
10 kHz
30
μ
s
30
μ
s
Max. number of retrofittable
digital inputs
Max.
pulse frequen-
cy
Min. pulse dura-
tion
Signal type
On
Off
max. 2 optional boards with 3 digi-
tal inputs each
1 Hz
300 ms
300 ms
Potential-free contact
Description
Switching out-
put
Ampacity at resis-
tive load
Contact life
a
a
The maximal ampacity of the contacts must not be exceeded.
OUT 1
Relay, normally
open contact
3 A at AC 250 V
150,000 switching cy-
cles
OUT 2
Relay, normally
open contact
OUT 3
Relay
changeover
contact
Summary of Contents for AQUIS touch S
Page 8: ...Inhalt ...
Page 20: ...10 Configuration 20 Chapter 10 12 Calibration timer page 46 ...
Page 68: ...13 Calibrating redox sensors 68 ...
Page 78: ...15 Calibrating CR conductivity sensors 78 ...
Page 88: ...16 Calibrating Ci conductivity sensors 88 ...
Page 96: ...17 Calibrating universal inputs 96 ...
Page 106: ...18 Calibrating O DO sensors 106 ...
Page 146: ...21 Annex 146 ...
Page 147: ......
Page 148: ......
Page 149: ......