21 Annex
144
(4.2 V). The voltage at sensor 2 (ecoLine O-DO) roughly corresponds to the
minimum value (5 V). For sensor 3 (ecoLine NTU), the voltage is not enough.
NOTE!
For the operation of JUMO ecoLine sensors, it is generally recommended to
use JUMO digiLine hubs and to generate the DC 5.3
ټ
V supply voltage within
the JUMO digiLine hub.
NOTE!
The voltage drop calculation shown here does not apply for the use of JUMO
tecLine sensors (types 20263x).