7
Drying Instructions
Dry the instrument immediately after final rinse. Use filtered
compressed air to dry internal areas. After drying proceed to
Inspection instructions.
INSPECTION AFTER CLEANING
• Inspect all instruments before sterilization or storage to
ensure the complete removal of soil from surfaces, tubes
and holes, moveable parts.
• If areas are difficult to inspect visually, check for blood by
immersing or flushing the instrument in a 3% hydrogen
peroxide solution. If bubbling is observed, blood is
present. Rinse instruments thoroughly after using
hydrogen peroxide solution.
• If soil is still present, reclean the instrument.
MAINTENANCE
Between uses, lubricate moving parts with a water-soluble
lubricant in accordance with the manufacturer’s instructions.
INSPECTION AND FUNCTIONAL TESTING
• Visually inspect the instrument and check for damage
and wear.
• Cutting edges should be free of nicks and have a
continuous edge.
• Jaws and teeth should align properly.
• Moveable parts should have smooth movement without
excessive play.
• Locking mechanisms should fasten securely and close
easily.
• Long, thin instruments should be free of bending
and distortion.
PACKAGING
• For instruments that are provided in sets, instrument
trays may be used for containment during sterilization.
• Package trays/instruments with a barrier wrap material
in accordance with local procedures, using standardized
wrapping techniques such as those described in ANSI/
AAMI ST46-1993.
STERILIZATION
• Use a validated, properly maintained and calibrated
steam sterilizer.
Summary of Contents for DePuy Mitek GRAFT PREPARATION SYSTEM
Page 1: ...GRAFT PREPARATION SYSTEM P N 106674 Rev F 01 11...
Page 2: ...1...
Page 72: ...71...
Page 73: ...72 PEI DePuy Mitek 1 1 2 3 4 5...
Page 74: ...73 6 7 1 2 DePuy Mitek...
Page 75: ...74...
Page 76: ...75 INSTRUMENTS pH 2 4 pH 3 60 C 20 1 82 2 C 5 95 C 10 30 C...
Page 77: ...76 38 49 C 1...
Page 78: ...77 3 ST46 1993 ANSI AAMI...
Page 106: ...105 DePuy MITEK...
Page 107: ...106 PEEK PEEK PEI DePuy Mitek 1...
Page 108: ...107 1 2 3 4 5 6 7 1...
Page 109: ...108 2 DePuy Mitek...
Page 110: ...109 pH 2 4 pH 3 60 C 20 1 82 2 C 5 95 C...
Page 111: ...110 10 30 C 1 38 49 C...
Page 112: ...111 3...
Page 113: ...112 ANSI AAMI ST46 1993 132 134 C 4 134 137 C 3...
Page 127: ...126 PEEK PEEK PEI DePuy Mitek 1...
Page 128: ...127 1 2 3 4 5 6 7 1 2 DePuy Mitek...
Page 129: ...128 pH 2 4 pH 60 C 3 20 82 2 C 1 95 C 5 10 30 C...
Page 130: ...129 38 49 C 1 3 ANSI AAMI ST46 1993...
Page 131: ...130 132 134 C 4 134 137 C 3 DePuy MITEK Tensioning Clamp Graft Clamp Suture Clamp...
Page 133: ...132 6 7 1 2 DePuy Mitek box lock flexible shaft washer disinfector...
Page 135: ...134 3 ANSI AAMI ST46 1993 132 134 C 4 134 137 C 3...
Page 136: ...135 DePuy Mitek...
Page 137: ...136 PEEK PEEK PEI PPSU DePuy Mitek 1 1 2 3 4...
Page 138: ...137 5 6 7 1 2 DePuy Mitek...
Page 139: ...138 pH 2 4 pH 60 C 3 20 82 2 C 1 95 C 5 10 30 C...
Page 140: ...139 38 49 C 1 3...
Page 141: ...140 ANSI AAMI ST46 1993 132 134 C 4 134 137 C 3...
Page 142: ...141...
Page 143: ...142...
Page 144: ...143...