5
AUTOMATED CLEANING
Instrument types with such complex design features as
cannulations, lumens, (i.e. tubes), holes, hinged joints,
box locks, spring-loaded retractable features, or flexible
shafts; must be manually pre-cleaned prior to automated
cleaning to improve the removal of adherent soil.
Manual Pre-Cleaning Instructions
• Flush internal areas with a warm detergent solution.
Pay close attention to lumens, cannulations, holes,
threads, crevices, seams, and any hard to reach areas.
While flushing, actuate any moveable mechanisms,
(hinged joints, box locks, or spring-loaded features),
to free trapped blood and debris. If the components of
the instrument can be retracted, retract or open the part
while flushing the area.
• For instruments with flexible shafts, flex the instrument
while flushing.
• For instruments with lumens, brush the internal surfaces
of the lumen.
Loading Instructions
• Load instruments into washer disinfector so that hinges
are open and cannulations and holes can drain.
• Place heavier instruments on the bottom of containers.
Do not place heavy instruments on top of delicate
instruments.
• For instruments with concave surfaces, such as curettes,
place instrument with the concave surface facing
downward to facilitate draining.
Automated Cleaning Instructions
• Clean in a validated washer disinfector using the
“INSTRUMENTS” cycle and a pH neutral cleaning
agent intended for use in automated cleaning. Effective
automated cleaning can be achieved using the following
parameters: two cold water pre-washes (minimum hold of
two minutes each); a hot water enzyme wash (minimum
hold of 4 minutes); followed by a hot water pH neutral
detergent wash (minimum hold 3 minutes at 60°C); a hot
water rinse (minimum time 20 seconds); a thermal rinse
(minimum hold of 1 minute at 82.2°C) and dry (minimum
hold 5 minutes at 95°C).
Summary of Contents for DePuy Mitek GRAFT PREPARATION SYSTEM
Page 1: ...GRAFT PREPARATION SYSTEM P N 106674 Rev F 01 11...
Page 2: ...1...
Page 72: ...71...
Page 73: ...72 PEI DePuy Mitek 1 1 2 3 4 5...
Page 74: ...73 6 7 1 2 DePuy Mitek...
Page 75: ...74...
Page 76: ...75 INSTRUMENTS pH 2 4 pH 3 60 C 20 1 82 2 C 5 95 C 10 30 C...
Page 77: ...76 38 49 C 1...
Page 78: ...77 3 ST46 1993 ANSI AAMI...
Page 106: ...105 DePuy MITEK...
Page 107: ...106 PEEK PEEK PEI DePuy Mitek 1...
Page 108: ...107 1 2 3 4 5 6 7 1...
Page 109: ...108 2 DePuy Mitek...
Page 110: ...109 pH 2 4 pH 3 60 C 20 1 82 2 C 5 95 C...
Page 111: ...110 10 30 C 1 38 49 C...
Page 112: ...111 3...
Page 113: ...112 ANSI AAMI ST46 1993 132 134 C 4 134 137 C 3...
Page 127: ...126 PEEK PEEK PEI DePuy Mitek 1...
Page 128: ...127 1 2 3 4 5 6 7 1 2 DePuy Mitek...
Page 129: ...128 pH 2 4 pH 60 C 3 20 82 2 C 1 95 C 5 10 30 C...
Page 130: ...129 38 49 C 1 3 ANSI AAMI ST46 1993...
Page 131: ...130 132 134 C 4 134 137 C 3 DePuy MITEK Tensioning Clamp Graft Clamp Suture Clamp...
Page 133: ...132 6 7 1 2 DePuy Mitek box lock flexible shaft washer disinfector...
Page 135: ...134 3 ANSI AAMI ST46 1993 132 134 C 4 134 137 C 3...
Page 136: ...135 DePuy Mitek...
Page 137: ...136 PEEK PEEK PEI PPSU DePuy Mitek 1 1 2 3 4...
Page 138: ...137 5 6 7 1 2 DePuy Mitek...
Page 139: ...138 pH 2 4 pH 60 C 3 20 82 2 C 1 95 C 5 10 30 C...
Page 140: ...139 38 49 C 1 3...
Page 141: ...140 ANSI AAMI ST46 1993 132 134 C 4 134 137 C 3...
Page 142: ...141...
Page 143: ...142...
Page 144: ...143...