Troubleshooting
60
DePuy Synthes
Battery Power Line II Instructions for Use
General
Problem Possible
causes
Solution
Drive unit does not start
No battery in the drive unit
Insert charged battery
Battery is discharged
Charge or replace battery
Battery is defect
Replace battery
Drive unit is defect
Send drive unit to DePuy Synthes
Service Center
Drive unit did not cool down after
sterilization
Allow to cool to room temperature
Mode switch is set to "lock" (off
position)
Set mode switch to ON/FWD/REV
No electrical contact between drive
unit and battery casing
Reinsert or change battery casing
Drive unit lacks power
Battery is not fully charged or past
life cycle
Charge or replace battery
Wrong attachment used (e.g. drill-
ing speed vs. reaming speed)
Change attachment
Drive unit has not been properly
serviced
Send drive unit to DePuy Synthes
Service Center
Attachments have not been prop-
erly serviced
Send attachments to DePuy Syn-
thes Service Center
Drive unit suddenly stops
Drive unit has overheated
Allow to cool to room temperature
Battery is empty/discharged
Charge or replace battery
Drive unit is defective
Send drive unit to DePuy Synthes
Service Center
Drive unit continues to run after
releasing trigger
Trigger is jammed by residue
Immediately turn mode switch to
"lock" (off position) or remove the
battery casing
Precaution:
Clean and lubricate
trigger according to care and
maintenance guidelines
Drive unit is defective
Immediately turn mode switch to
"lock" (off position) or remove the
battery casing Send drive unit to
DePuy Synthes Service Center
Drive unit or attachment becomes
excessively hot
Drive unit or attachment is used
outside the specification
Allow drive unit or attachment to
cool. (see Duty cycles on page 64)
The cutting tool is blunt
Replace the cutting tool
Visible physical damage on items
Battery was accidentally reproc-
essed
Replace battery. Send damaged
battery to DePuy Synthes Service
Center
Drive unit, attachment, battery cas-
ing, sterile cover was dropped
Replace damaged items. Send
damaged items to DePuy Synthes
Service Center
Battery is faulty
Staff negligence
Replace battery and send the bat-
tery to DePuy Synthes Service
Center
Battery casing jams when insert-
ing or removing from drive unit
Coupling mechanism has not been
lubricated
Clean and lubricate according to
care and maintenance guidelines
Coupling mechanism is damaged
Send damaged item to DePuy
Synthes Service Center
Battery casing lid is difficult to
open and close
Sealing ring has not been lubri-
cated
Clean and lubricate according to
care and maintenance guidelines