Operating Instructions
Battery Pack (battery casing with inserted battery)
16
DePuy Synthes
Battery Power Line II Instructions for Use
Scrubbed person
Close the battery casing (Figs. 7a and 7b).
Note:
Ensure that both battery casing locks engage and
that the lid of the battery casing is closed properly.
Always ensure that the lid of the battery casing is
totally closed before using the system.
Precaution:
Do not contact the non-sterile battery or the
inside of the battery casing in order to avoid con-
tamination. Should the scrubbed person come into
contact with either the non-sterile battery or the in-
side of the battery casing, they must be scrubbed
again. Should the battery casing be contaminated it
must be cleaned and resterilized before being used
in the operating room.
Notes:
–
Normally, one fully charged battery has sufficient
capacity for an entire operation. As a precaution,
a second battery pack (battery casing with in-
serted battery) should be kept ready, so that the
battery pack can be quickly exchanged under
sterile conditions during surgery if necessary.
–
Never open a battery casing intraoperatively to
insert a new battery. Always replace the whole
battery pack by another battery pack prepared
before the start of the surgery.
Insert the battery pack into the drive unit, ensuring
the contacts on the battery pack align with the con-
tacts in the recess of the drive unit (Fig. 8). Press
firmly to ensure the battery pack is engaged cor-
rectly, and check by pulling lightly downward on the
battery pack.
Notes:
–
For safety reasons, the battery pack can only be
inserted fully when it is in the correct orientation.
–
To prevent injuries, the mode switch of the drive
unit should always be in the off position when in-
serting or removing the battery pack.
–
Installing the battery pack just before use pre-
vents unwanted discharge of battery capacity.
Figure 7a
Figure 7b
Figure 8
Scrubbed person
530.705
530.690
Scrubbed person
Scrubbed person
Summary of Contents for BPL II
Page 2: ......
Page 4: ...2 DePuy Synthes Battery Power Line II Instructions for Use...
Page 75: ......