6. Visually inspect device
Inspect the cannulations, sliding
sleeves, attachment
release rings, etc. for visible soil. Repeat steps 1-6
until no visible
soil remains.
7. Final rinse with de-ionized/purified water
Final rinse with de-ionized or purified water for a
minimum
of 2 minutes (Fig. 7).
Figure 7
8. Dry
Dry device using a clean, soft lint-free cloth or clean
compressed air (Fig. 8).
Figure 8
Battery Power Line II Instructions for Use
DePuy Synthes
43