Operating Instructions
Battery Pack (battery casing with inserted battery)
Synthes non-sterile batteries and advanced charging
technology optimize intraoperative battery capacity,
maximize battery lifespan and shorten turnaround
time. One Universal Battery Charger II (05.001.204)
for multiple Synthes battery-driven systems simplifies
the charging process. Simple aseptic technique pre-
serves the sterile field when assembling the battery
pack.
Instruments
530.630
Battery for Battery Power Line II
530.660
Sterile Cover for Battery Power Line II
530.690
Battery Casing for Battery Power Line II
Assembling and inserting the battery pack
Scrubbed person
Open the lid of the battery casing as shown in Fig. 1.
Ensure the lid of the battery casing is fully open
(Fig. 2). Ensure that the lid of the battery casing is
facing towards the scrubbed person (Fig. 3).
Position the sterile cover securely on top of the bat-
tery casing (Fig. 4).
Notes:
–
The sterile cover helps to guide the battery into
the battery casing and prevents contamination of
the sterile casing by the non-sterile battery.
–
Sterilize the sterile cover after each use to en-
sure aseptic conditions when inserting the un-
sterile battery into the sterile battery casing.
–
If the non-sterile battery contacts the outside of
the battery casing, the battery casing must be
cleaned and resterilized before being used in the
operating room.
–
Do not insert the non-sterile battery into the bat-
tery casing whilst a handpiece is attached.
Figure 1
Figure 2
Figure 3
530.690
Scrubbed person
Scrubbed person
Scrubbed person
Scrubbed person
530.660
530.690
Figure 4
14
DePuy Synthes
Battery Power Line II Instructions for Use