Battery Power Line II Instructions for Use
DePuy Synthes
49
Battery Reamer/Drill II (530.705)
The following individual parts must be lubricated with
1 drop of Synthes Special Oil (519.970):
1
Attachment release ring (Figs. 1a and 1b)
2
Trigger shaft
3
Rear end of the cannulation
Turn the attachment release ring clockwise and
insert 1 drop of Synthes Special Oil (519.970) as
shown in figure 1a. Then turn the release ring sev-
eral times.
Insert 1 drop of Synthes Special Oil (519.970) in the
gap between the seal ring and shaft (Fig. 1b). Insert
the battery pack and run handpiece to ensure oil is
evenly distributed.
Figure 1a
Battery casing release buttons from the inside, after
which the buttons should be pressed several times
(Fig. 2)
Figure 2
Figure 1b