Care and Maintenance
Manual Cleaning Instructions
42
DePuy Synthes
Battery Power Line II Instructions for Use
4. Clean with detergent
Clean the device manually under running warm wa-
ter using an enzymatic cleaner or detergent for a
minimum of 5 minutes. Manipulate all moving parts
under running water. Use a soft-bristled brush and/or
soft lint-free cloth to remove all visible soil and debris
(Figs. 3 and 4). Follow the enzymatic cleaner or de-
tergent manufacturer’s instructions for use for correct
temperature, water quality and concentration/dilution.
Note:
For the Quick Coupling for Pins Ø 3.2 mm (530.796)
the brush should only be inserted from the front.
5. Rinse with tap water
Rinse the device thoroughly using cool to lukewarm
running water for a minimum of 2 minutes.
Use a
syringe or pipette to flush lumens and channels.
Actuate joints, handles and other movable device
features in order to rinse thoroughly under running
water.
Figure 3
Figure 4: Quick Coupling for Kirschner Wires and for Pins
Ø 1.5 – 4.0 mm (530.791)