Operating Instructions
Attachments for Battery Reamer/Drill II
24
DePuy Synthes
Battery Power Line II Instructions for Use
Insert instrument
Open the chuck jaws by turning the key (510.191)
counterclockwise, or by manually turning the collar
(Fig. 1).
Insert the instrument shaft into the opened chuck.
Close the chuck manually by rotating the collar, keep-
ing the instrument shaft centered in the jaws. Tighten
the chuck by turning the key clockwise (Fig. 2).
Note:
To ensure secure fixation of the instrument, make
sure the toothed rims on the drill chuck and key are
not worn. Replace damaged or worn components.
Remove instrument
Turn the key counterclockwise to open the jaws.
Remove the instrument.
Figure 1
Figure 2