8. Automated cleaning cycle parameters
Note:
The washer/disinfector should fulfill requirements specified in ISO 15883.
Step
Duration (minimum)
Cleaning Instructions
Rinse
2 minutes
Cold tap water
Pre-wash
1 minute
Warm water (
≥
40°C); use detergent
Clean
2 minutes
Warm water (
≥
45°C); use detergent
Rinse
5 minutes
Rinse with de-ionized (DI) or purified water
(PURW)
Thermal processing
5 minutes
Hot DI water,
≥
93°C
Dry 40
minutes
≥
90°C
9. Inspect device
Remove all devices from the washing basket.
Inspect the cannulations, sliding sleeves, etc. for
visible soil. If necessary, repeat the manual pre-
cleaning/automated cleaning cycle. Confirm that all
parts are completely dry.
Precaution:
Mechanical cleaning is an additional stress for power
equipment, especially for seals and bearings. There-
fore, devices must be properly lubricated after auto-
mated cleaning. Furthermore, the device must be
serviced at least once per year as specified under
the section “Repair and Technical Services” on page
58.