13.14.3 Synchronization priority
An external synchronization is performed based
on different priorities where:
·
PrForFty 1: Modbus synchronFzatFon
For this, via the Modbus tool, set the
“Enable flag” (Addr: 822) or select the option
“Synchronization via Modbus” in the GridVis
®
software in the configuration area for the trailer
pointer.
·
PrForFty 2: SynchronFzatFon vFa dFiFtal Fnput 3
For this, set the Modbus parameter “FUNC_
SYNC_RECORD” (Addr. 30048, to value 4) or
select the option “Trailer pointer synchronization”
in the GridVis
®
software in the configuration area
for peripherals (digital input 3).
Note: Do
NOT
select the “Synchronization
via Modbus” option in the trailer pointer
configuration!
·Priority 3: Internal synchronization
Modbus
address
Function
Setting
range
820
Sets a trigger flag for the
trailer pointer synchronization
0 .. 1
821
Time base in seconds
60 .. 65535
822
Enable flag of the
Modbus trigger
0 .. 1
823
Capture time in seconds
0 .. 255
30048
Input configurations
0 .. 4*
Fig.
Table Modbus address for synchronFzatFon
* 0 = UNC_NONE;
1 = UNC_TARI ;
2 = UNC_SYNC_CLOCK_MIN;
3 = UNC_SYNC_CLOCK_H;
4 = UNC_SYNC_RECORD