PhaseStation 53100A User’s Manual V1.04 25-Apr-2022
54
In the example at left, two copies of
the same 3600-sample .TIM file have
been loaded. The
blue
trace was
rendered with the default
Bin
Threshold
of 4, while the shorter
magenta
trace was rendered with
Bin
Threshold
set to 2000. Enabling
Trace
→
Show xDEV error bars (
Ctrl-e
)
reveals the
blue
trace’s low
confidence at its longest tau.
Generally, as long as
Trace
→Clip xDEV traces by confidence
(
Ctrl-v
)
is enabled, it will
not be necessary to alter the
Bin Threshold
value to display xDEV trace data of good
statistical quality. The default bin threshold and clipping options will suffice for almost
all measurements, particularly those made at high data rates by instruments such as the
PhaseStation 53100A.
Figure 17: Bin threshold adjustment
Summary of Contents for PHASE STATION 53100A
Page 1: ...TM 53100A Phase Noise Test Set User s Manual Revision 1 04 April 25 2022...
Page 16: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 16...
Page 18: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 18...
Page 20: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 20...
Page 30: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 30...
Page 36: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 36...
Page 42: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 42...
Page 64: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 64...
Page 74: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 74...
Page 104: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 104...
Page 116: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 116...
Page 122: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 122...
Page 130: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 130...
Page 140: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 140...
Page 214: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 214...
Page 215: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 215 Appendix Legal and regulatory notices...