PhaseStation 53100A User’s Manual V1.04 25-Apr-2022
149
value with
AcqSetParam(),
or (if
skip_dialog
is false) the user’s selection of the
Cancel
button.
Once the measurement begins collecting data, the script’s
EventAcqTriggered()
event
handler (if any) will be called. Prior to this stage, the script’s
EventAcqSetup()
handler will
be called if one is defined. Because
AcqStartAcquisition()
returns immediately after
launching the acquisition thread, measurement-specific parameters such as
Input Freq
will not be available to the script until
EventAcqTriggered()
is called.
Optionally, a third parameter may be passed, consisting of an arbitrary numeric
user_value
which will be made available to subsequent event handlers.
AcqStopAcquisition()
Immediately stops any acquisition associated with the running script instance and calls
the script’s
EventAcqDone()
handler (if any). Unless a fatal error occurs, the plot(s) will not
be closed and the script will continue to run until
ScriptEnd()
is called.
AcqSetupEnd()
Applications that include
EventAcqSetup()
handlers must call
AcqSetupEnd()
when their setup
phase is complete. Until
AcqSetupEnd()
is called, the 53100A driver will remain in an
endless loop awaiting further setup commands from the script.
DisplayFontSize([Number points])
Sets or returns the size of the main display font size used in the TimeLab window. This is
the same value that is adjusted by the commands on the
Display
menu.
If no parameter is specified, the current font size is returned without alteration.
DisplayOverlayMode([Boolean overlay_mode])
Equivalent to the
Display
→Browse plots one at a time (
b
)
or
Display→Overlay all
loaded plots (
o
)
command, depending on the
overlay_mode
parameter.
The
overlay_mode
parameter is optional; if it is not provided by the script, the function
simply returns true in
Overlay
mode or false in
Browse
mode. If a new
overlay_mode
value is specified, the function returns the previous display mode.
DisplayShowNumTab([Boolean status])
Sets and/or retrieves the
status
of the flag controlled by
Display→Numeric table (
Ctrl-n
)
.
Refer to this command for more information.
To return the current flag value, call
DisplayShowNumTab()
with no argument.
Summary of Contents for PHASE STATION 53100A
Page 1: ...TM 53100A Phase Noise Test Set User s Manual Revision 1 04 April 25 2022...
Page 16: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 16...
Page 18: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 18...
Page 20: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 20...
Page 30: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 30...
Page 36: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 36...
Page 42: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 42...
Page 64: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 64...
Page 74: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 74...
Page 104: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 104...
Page 116: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 116...
Page 122: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 122...
Page 130: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 130...
Page 140: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 140...
Page 214: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 214...
Page 215: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 215 Appendix Legal and regulatory notices...