PhaseStation 53100A User’s Manual V1.04 25-Apr-2022
196
exit
See
quit.
help [<command>]
Returns a list of commands with brief usage information. If a command name is specified, only
the help text for that command is returned.
history
Returns a list of previously-entered commands.
logout
See
quit.
measurelinear
Computes the slope and intercept coefficients of the current phase-difference chart. After a
measurelinear
command has been issued, a subsequent
removelinear on
command will cause
the linear trendline to be removed from the chart data returned by
show phasediff.
If
measurelinear
is executed while
removelinear on
is in effect, the stored coefficients will be
updated for use by subsequent
show phasediff
commands. It is not necessary to use
removelinear off
before issuing a
measurelinear
command.
open <filename>
Opens an ASCII text file to which streaming phase-difference or frequency readings will be
written. If the specified file already exists, it will be overwritten. Typically the file is shared with
other applications that can access it in read-only mode while it’s being written by TSERVE.
The data written to the file may be ‘followed’ from within TimeLab by selecting
Acquire→Acquire from live ASCII file
.
You can also monitor the data written to the file with a
command such as
tail –f <filename>
from Cygwin. Finally, pressing the
Scroll Lock
key will cause
TSERVE to display the lines written to the file on the stdout device (i.e., the DOS console itself)
at a maximum rate of 10 readings per second.
Example:
tserve64 /data_port:0 /cmd_port:0 /start /format:P /file:%HOMEPATH%\Documents\mydata.txt
The command above writes a continuous stream of phase-difference readings to the file
mydata.txt
in the current user’s Documents folder. No TCP/IP ports are opened for either data
streaming or command input. You can then use TimeLab or another application to ‘follow’ the
file. Even if the acquisition fails or otherwise encounters problems, you can use the appropriate
File→Import
option to bring the data into TimeLab at any time.
The output file may be flushed and closed at any time by issuing a
close
command or by
open
ing
a different file. In addition to the
open
command, a stream output file can also be opened at
TSERVE startup time by specifying the
/file: <filename>
command-line option.
Summary of Contents for PHASE STATION 53100A
Page 1: ...TM 53100A Phase Noise Test Set User s Manual Revision 1 04 April 25 2022...
Page 16: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 16...
Page 18: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 18...
Page 20: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 20...
Page 30: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 30...
Page 36: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 36...
Page 42: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 42...
Page 64: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 64...
Page 74: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 74...
Page 104: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 104...
Page 116: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 116...
Page 122: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 122...
Page 130: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 130...
Page 140: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 140...
Page 214: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 214...
Page 215: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 215 Appendix Legal and regulatory notices...