PhaseStation 53100A User’s Manual V1.04 25-Apr-2022
197
pause <chart>
pause phasediff
and
pause freq
will inhibit updates to the phase difference and frequency
difference charts, respectively. Updates may be resumed with
resume phasediff
or
resume
freq
. (The inconsistent chart nomenclature is required for compatibility with the TSC 51XXA
command language.)
prompt <on|off>
The
prompt off
and
prompt on
commands will disable and enable the Telnet command prompt,
respectively. The prompt is enabled by default.
prompt off
is equivalent to the
/noprompt
command-line option.
quit
Closes the Telnet connection. The TSERVE server will continue running, as will any
measurement in progress.
removelinear <on|off>
When
removelinear on
is in effect, the linear trendline most recently measured by the
measurelinear
command will be removed from subsequent phase-difference charts returned by
the
show phasediff
command. Use
removelinear off
to disable linear-trend subtraction.
reset
restorefactorydefaults
When the Telnet server thread is created by TSERVE at startup time, copies of the default values
for the
phaserate
,
tau0
,
referencefreq
,
inputfreq, prompt
,
dateformat
,
timeformat
, and
title
properties are associated with the server instance, taking into account any changes made to the
program defaults at the TSERVE command line. These properties subsequently apply to any
measurements initiated by either Telnet clients or by a
/start
parameter on the TSERVE
command line. Clients may alter these server property values by issuing the appropriate
set
commands. Changes made to the server property values will apply to acquisitions that are
subsequently initiated by that client or any others.
Issuing a
reset
or
restorefactorydefaults
command will immediately stop any measurement in
progress. TSERVE will continue running and any clients will remain connected, but the server
default properties will be restored to the program default or command-line values, just as if
TSERVE had been terminated and restarted. The
reset
and
restorefactorydefaults
commands
behave identically.
resume <chart>
See
pause
.
Summary of Contents for PHASE STATION 53100A
Page 1: ...TM 53100A Phase Noise Test Set User s Manual Revision 1 04 April 25 2022...
Page 16: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 16...
Page 18: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 18...
Page 20: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 20...
Page 30: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 30...
Page 36: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 36...
Page 42: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 42...
Page 64: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 64...
Page 74: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 74...
Page 104: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 104...
Page 116: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 116...
Page 122: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 122...
Page 130: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 130...
Page 140: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 140...
Page 214: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 214...
Page 215: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 215 Appendix Legal and regulatory notices...