
PRS-330 Series
APPENDIX B
54
CONFIGURE Tables
CONFigure:RTD <P100C, P100F, P1000C,P1000F>
CONFigure:RTD? Returns which RTD table is selected P100C, P100F, P1000C, or P1000F
CONFigure:TABLe:SELect <number> Selects Table to be used, <number> is 0 to 9. 0 is no
table with display shown in ohms and 1 , 2, 3 and 4 are prede
fi
ned and 5 to 9 are user de
fi
ned tables
CONFigure:TABLe:SELect? Returns table number that is selected
CONFigure:TABLe:NAME <name> Sets Table name, <name> is ASCII up to 20 characters
such as “PT100”
CONFigure:TABLe:NAME? Returns the number, 1 to 4 ,of currently selected table
CONFigure:TABLe:ERASE Erases all rows in the table
CONFigure:TABLe:ADD <user_value,resistance_value> Adds a row to the end of table show-
ing user value and resistance value in ohms.
example CONF:TABL:ADD 20, 95 means that if the table is selected and the operator enters 20
from the front panel the output will be set to 90
Ω
.
CONFigure:TABLe:ADD? Returns the last entry in the table
CONFigure:TABLe:DISPlay? Returns all data in the table
Example of information returned
PT100
20, 95
CONFigure:TABLe:UNIT <Unit> Set units for tables up to 8 ASCII characters
As an example “C” could be send for Celsius or “F” for Fahrenheit
CONFigure:TABLe:UNIT? Returns ASCII string representing the units
Summary of Contents for PRS-330 series
Page 2: ...TEL 516 334 5959 800 899 8438 FAX 516 334 5988 www ietlabs com IET LABS INC...
Page 12: ...PRS 330 Series v Table of Contents Page left intensionally blank...
Page 18: ...SPECIFICATIONS PRS 330 Series 6 Page left intensionally blank...
Page 34: ...OPERATION PRS 330 Series 22 Page left intensionally blank...
Page 36: ...USB INTERFACE PRS 330 Series 24 Page left intensionally blank...
Page 44: ...PRS 330 Series ETHERNET INTERFACE OPTION 32...
Page 48: ...PROGRAMMING PRS 330 Series 36 Page left intensionally blank...
Page 67: ...PRS 330 Series APPENDIX B 55...