SPECIFICATIONS
PRS-330 Series
5
FIGURE 2.1 Typical OPERATING GUIDE Affi xed to Unit
(Please see label affi xed to your unit)
PRS-330 PROGRAMMABLE DECADE RESISTANCE SUBSTITUTER
CONSULT INSTRUCTION MANUAL FOR PROPER INSTRUMENT OPERATION
MODEL: PRS-330 SN: B2-18431158
PRS-330 Nov. 2018
WARNING
Observe all safety rules when working with high voltages or line voltages. Connect the (GND) terminal to earth ground in order to maintain the case at a safe voltage.
Whenever hazardous voltages (>45 V) are used, take all measures to avoid accidental contact with any live components: a) Use maximum insulation and minimize the use
of bare conductors. b) Remove power when adjusting switches. c) Post warning signs and keep personnel safely away.
IET LABS, INC.
www.ietlabs.com
CAGE CODE: 62015
• Long Island, New York
• Tel: (516) 334-5959 • Email: [email protected]
Accuracy:
±(0.007% + 1 m
Ω
) 2 and 4 Terminal at 23°C ±5°C
Minimum setting:
0.100 000
Ω
Resolution:
1
μΩ
or 6 digits
Stability:
±50 ppm/year
Range:
0.1
Ω
- 20 M
Ω
Thermal emf:
<15
μ
V
Maximum Load:
2 A, 200 V (peak), 0.5 W whichever applies
fi
rst
Resistors:
Precision wire-wound and metal foil
RTD Tables:
PT-100 and PT-1000 tables for both Fahrenheit and Celsius are programmed
into memory locations.
9 RTD tables can be entered into memory to allow user selection of temperature
and the correct value of resistance will automatically be programmed.
Adjustment:
Automatic adjustment procedure utilizing a high precision DMM eliminates the
requirement for manual trimming of resistors.
Relays
1
:
Silver Alloy contacts, expected life of 10
8
cycles
Terminals
:
Four low-emf, gold-plated, tellurium-cop per 5-way bind ing posts are used for
HI
and
LO
ter mi nal pairs for
CUR RENT
and
SENSE
.
GND
bind ing post is con nect ed
to the case, to the chassis ground and to the earth ground.
Switching time:
<1 second per change
Power requirements:
90 - 264 Vac , 47 - 63 Hz. 30 W. Max.
Remote Control:
USB:
Standard USB Type B connector on rear panel, uses MCP2200 driver
Remote Control Options:
IEEE:
GPIB standard 24 pin connector, conforms to IEEE-488.2; SCPI 1994.0
command set; Hardware or software con
fi
gurable addressing range of 1 to 30.
E t h e r n e t :
I E E E 8 0 2 . 3 c o m p l i a n t , S p e e d s 1 0 B a s e T
(10 Mb/s) and 100 BaseT (100 Mb/s), IP Address Static or DHCP, Factory setting
192.168.0.254 static
1
Note: Warranty covers relays up to expected life
Summary of Contents for PRS-330 series
Page 2: ...TEL 516 334 5959 800 899 8438 FAX 516 334 5988 www ietlabs com IET LABS INC...
Page 12: ...PRS 330 Series v Table of Contents Page left intensionally blank...
Page 18: ...SPECIFICATIONS PRS 330 Series 6 Page left intensionally blank...
Page 34: ...OPERATION PRS 330 Series 22 Page left intensionally blank...
Page 36: ...USB INTERFACE PRS 330 Series 24 Page left intensionally blank...
Page 44: ...PRS 330 Series ETHERNET INTERFACE OPTION 32...
Page 48: ...PROGRAMMING PRS 330 Series 36 Page left intensionally blank...
Page 67: ...PRS 330 Series APPENDIX B 55...