9.3.4 Adjust PRS
Clicking on the Adjust PRS key will step the
operator through the PRS-330 adjustment
procedure. The procedure consists of measuring
all 43 internal resistors using the DMM and
storing the actual resistance values into a table.
These resistance values are used in the algorithm
to create the programmed resistance value. See
section 8.1.4 for more information.
9.3.5 PRS-DMM Speed
There are three selections for PRS-DMM Speed.
These selection affect the accuracy from the
target value and time to make the measurements.
Slow measurements are made with 7 ½ digits of
resolution, and measurements and adjustments
are made until the accuracy is better than ±20
ppm.
Medium measurements are made with 6 ½ digits
of resolution and measurements and adjustments
are made until the accuracy is better than 40 ppm.
Fast measurements are made with 6 ½ digits
of resolution. There are no adjustments made
in fast mode. The value from the DMM is just
displayed.
The uncertainty of the measurement is also
displayed. The uncertainty is based upon the
speci
fi
cations given in the manual for the Fluke
8508A or Agilent 3458A. The uncertainty is
based upon either 6 ½ or 7 ½ digits of resolution,
absolute uncertainty for 365 days and a 95%
con
fi
dence level.
Summary of Contents for PRS-330 series
Page 2: ...TEL 516 334 5959 800 899 8438 FAX 516 334 5988 www ietlabs com IET LABS INC...
Page 12: ...PRS 330 Series v Table of Contents Page left intensionally blank...
Page 18: ...SPECIFICATIONS PRS 330 Series 6 Page left intensionally blank...
Page 34: ...OPERATION PRS 330 Series 22 Page left intensionally blank...
Page 36: ...USB INTERFACE PRS 330 Series 24 Page left intensionally blank...
Page 44: ...PRS 330 Series ETHERNET INTERFACE OPTION 32...
Page 48: ...PROGRAMMING PRS 330 Series 36 Page left intensionally blank...
Page 67: ...PRS 330 Series APPENDIX B 55...