MAINTENANCE
PRS-330 Series
38
the memory of the PRS-330. The measured
values of the resistors are used in an algorithm
to determine which resistors are used to create
the programmed resistance value.
In addition to the 43 resistors the residual
internal lead resistance is also measured at time
of manufacture and entered into memory. The
internal lead resistance does not need to be
measured again unless repair has been performed
to the PRS-330.
Below is a list of the 43 resistors and their
nominal resistance value. These resistors should
be within ±1% of the nominal value for R1 to R5
and ±0.1% of the nominal value for R6 and higher
to ensure proper operation of the PRS-330. If the
measured value of the resistor is outside of these
speci
fi
cations service is required.
Resistance
Number
Nominal Resistance
Value
R1
0.17
Ω
R2
0.35
Ω
R3
0.7
Ω
R4
1.37
Ω
R5
2.66
Ω
R6
5.15
Ω
R7
10.2
Ω
R8
20.0
Ω
R9
38.8
Ω
R10
75.6
Ω
R11
148.0
Ω
R12
287.0
Ω
R13
570.0
Ω
R14
1.13 k
Ω
R15
2.23 k
Ω
R16
4.4 k
Ω
R17
8.7 k
Ω
R18
17.0 k
Ω
R19
33.3 k
Ω
R20
66.0 k
Ω
R21
130.0 k
Ω
R22
258.0 k
Ω
R23
506.0 k
Ω
R24
1.0 M
Ω
R25
2.0 M
Ω
R26
3.99 M
Ω
R27
7.8 M
Ω
R28
15.0 M
Ω
R29
30.0 M
Ω
R30
60.0 M
Ω
R31
120.0 M
Ω
R32
3.55 k
Ω
R33
7.39 k
Ω
R34
16.4 k
Ω
R35
33.8 k
Ω
Summary of Contents for PRS-330 series
Page 2: ...TEL 516 334 5959 800 899 8438 FAX 516 334 5988 www ietlabs com IET LABS INC...
Page 12: ...PRS 330 Series v Table of Contents Page left intensionally blank...
Page 18: ...SPECIFICATIONS PRS 330 Series 6 Page left intensionally blank...
Page 34: ...OPERATION PRS 330 Series 22 Page left intensionally blank...
Page 36: ...USB INTERFACE PRS 330 Series 24 Page left intensionally blank...
Page 44: ...PRS 330 Series ETHERNET INTERFACE OPTION 32...
Page 48: ...PROGRAMMING PRS 330 Series 36 Page left intensionally blank...
Page 67: ...PRS 330 Series APPENDIX B 55...