INTRODUCTION
PRS-330 Series
2
direct resistance substitution as well as RTD
(Resistance Temperature Detector) simulation. It
allows for higher precision and tight tolerances.
The PRS-330 is a precision resistance source
with excellent characteristics of accuracy,
stability, temperature coef
fi
cient, and power
coef
fi
cient. It is direct reading without
the need to subtract “zero resistance.”
The PRS-330 offers the option of true
4-wire Kelvin measurement for calibration
applications and 2-wire mode for use
in series as a substitute resistor and for
RTD applications. The front panel display
indicates the mode selected: 2-wire or 4-wire.
There are 4 pre-programmed RTD
resistance tables, and others may be added
by the user. The controller allows for
other convenient programmed features.
The unique design makes adjustment of the
PRS-330 for calibration semiautomatic. An
internal calibration routine guides the user to
perform measurements using only a precision
ohmmeter and enter the measured values into
the PRS-330. No other standards are required.
IET Labs offers the PRS-DMM Self Adjusting
Software which works with the PRS-330 and
Fluke 8508A or Keysight 3458A to automate
this adjustment process and eliminate the
need for a technician to manual enter the
resistance values manually into the PRS-
330 See Chapter 9 for more information
Summary of Contents for PRS-330 series
Page 2: ...TEL 516 334 5959 800 899 8438 FAX 516 334 5988 www ietlabs com IET LABS INC...
Page 12: ...PRS 330 Series v Table of Contents Page left intensionally blank...
Page 18: ...SPECIFICATIONS PRS 330 Series 6 Page left intensionally blank...
Page 34: ...OPERATION PRS 330 Series 22 Page left intensionally blank...
Page 36: ...USB INTERFACE PRS 330 Series 24 Page left intensionally blank...
Page 44: ...PRS 330 Series ETHERNET INTERFACE OPTION 32...
Page 48: ...PROGRAMMING PRS 330 Series 36 Page left intensionally blank...
Page 67: ...PRS 330 Series APPENDIX B 55...