OPERATION
PRS-330 Series
16
Figure 3-3 Resistance Settings Menu
There are 5 selections available in the
RESISTANCE SETTINGS
menu.
Highlight the selection and press
ENTER
to
change the selection.
Connection Mode:
allows the user to change
between 2-Wire and 4-Wire connections.
In 2-Wire the shorting links should be connected
between the two LO terminals, and between the
two HI terminals.
In 2-Wire mode the internal trace resistance
inside the PRS-330 is automatically compensated
for, so no zero subtraction is required.
In 4-Wire mode shorting links should be removed
from the binding posts. This mode was designed
to be used with DMMs, other meters with 4
terminal (Kelvin connections) and temperature
controllers and indicator using 4 terminal RTDs.
Terminal Selection:
allows selection of either
rear binding posts or front binding posts to be
connected to the source resistance.
There is no signi
fi
cant difference in accuracy
Switch Mode:
Allows selection of Default,
Open or Short modes. When the resistance or
temperature value is changed via the front panel
or when a SOURCE:DATA command is sent
via the remote interfaces, the intermediate value
between previous value and the new value will
either be a SHORT, OPEN or undetermined in
default switching case.
Standard Values and RTD and Other Tables
Settings
have separate menus which are
described in the sections below.
3.4.5 Standard Values Menu
There are 4 selections available in the
STANDARD
VALUES
menu. The
STANDARD VALUES
menu is only applicable to resistance values and
not RTD or Tables.
The selection of the standard values controls how
the
INC
and
DEC
softkeys work on the
Home
screen see Figure 3-1.
When a value is entered into one of the 4
selections such as Percent Value: shown in Figure
3-4 the
INC
and
DEC
keys will then in this
example increment or decrement the resistance
value by 10% each time the key is pressed.
An
*
before the selection indicates which is
active. The last selection to which a value was
entered becomes the active selection.
3.4.5.1 Set Values
The
Set Value:
selection allows the resistance
value to be changed by a speci
fi
c value in
Ω
s each
time the
INC
, increment, and
DEC
, decrement,
key is pressed.
Highlight
Set Value:
and press
ENTER
to
change the set value.
When a
Set Value:
is selected on the PRS-
330 pressing the
INC
and
DEC
softkeys will
increment the resistance value to the next
resistance value plus or minus the
Set Value
.
3.4.5.2 Standard Value
Summary of Contents for PRS-330 series
Page 2: ...TEL 516 334 5959 800 899 8438 FAX 516 334 5988 www ietlabs com IET LABS INC...
Page 12: ...PRS 330 Series v Table of Contents Page left intensionally blank...
Page 18: ...SPECIFICATIONS PRS 330 Series 6 Page left intensionally blank...
Page 34: ...OPERATION PRS 330 Series 22 Page left intensionally blank...
Page 36: ...USB INTERFACE PRS 330 Series 24 Page left intensionally blank...
Page 44: ...PRS 330 Series ETHERNET INTERFACE OPTION 32...
Page 48: ...PROGRAMMING PRS 330 Series 36 Page left intensionally blank...
Page 67: ...PRS 330 Series APPENDIX B 55...