PRS-DMM SOFTWARE
PRS-330 Series
45
will be shown in the white box along with a
Success indication.
Click Exit key to return to Setting Screen once
PRS has been detected.
Figure 9.3 PRS Settings Screen
9.3.3 Meter Settings
Meter Setting screen allows the operator to select
which DMM will be used either the Fluke 8508A
or the Keysight 3458A. The GPIB address of the
DMM can also be set in this menu. Make sure
the DMM is turned on and connected to the PC.
Figure 9.4 Meter Screen
Select the GPIB address and then click the
Initialize key. The IDN string will be shown in
the white box along with a Success indication.
The PRS-DMM software will configure the
DMM for resistance measurements, appropriate
resolution 7 ½ digits for slow, 6 ½ digits for
medium and fast, 4 wire measurements and
autorange ON.
The 8508A will also be configured for Low
Current, Fast AD Integration OFF, and Analog
Filter OFF.
If additional commands need to be sent to the
DMM, click on the Con
fi
gure Key and enter the
commands in the text box.
Once the commands have been entered press the
Con
fi
gure key. Any commands will be send to
the DMM during initialization of the DMM.
Figure 9.5 Meter Con
fi
gure Screen
Click Exit key to return to Setting Screen once
DMM has been detected.
Summary of Contents for PRS-330 series
Page 2: ...TEL 516 334 5959 800 899 8438 FAX 516 334 5988 www ietlabs com IET LABS INC...
Page 12: ...PRS 330 Series v Table of Contents Page left intensionally blank...
Page 18: ...SPECIFICATIONS PRS 330 Series 6 Page left intensionally blank...
Page 34: ...OPERATION PRS 330 Series 22 Page left intensionally blank...
Page 36: ...USB INTERFACE PRS 330 Series 24 Page left intensionally blank...
Page 44: ...PRS 330 Series ETHERNET INTERFACE OPTION 32...
Page 48: ...PROGRAMMING PRS 330 Series 36 Page left intensionally blank...
Page 67: ...PRS 330 Series APPENDIX B 55...