PRS-330 Series
OPERATION
17
The
Standard Value:
allows the PRS-330 to be
used as an automated resistance carousel with
built-in EIA “preferred value” resistance tables
of 1% (E96), 5% (E24), 10% (E12) increments
or RATIOMETRIC user speci
fi
ed increment.
Figure 3-4 Standard Values Menu
Resistors are available in a number of standard
ranges, often called ‘preferred values’ These
ranges, or series, are set out by the Electronic
Industries Association (EIA), and common
values are E12, E24, E48, and E96. The number
after the ‘E’ denotes the number of resistance
values the series contains per decade.
The series are logarithmic and are derived from
the resistor tolerance; resistors with a tighter
tolerance can have more values in the series
that won’t overlap one another. The series are
sometimes referred to by the tolerance, the two
being related as follows:
E12: 10% tolerance
E24: 5% tolerance
E48: 2% tolerance
E96: 1% tolerance
When a
Standard Value:
is selected on the
PRS-330 pressing the
INC
and
DEC
softkeys
will increment the resistance value to the next
“preferred” resistance value.
RATIOMETRIC allows a custom tolerance value
to be entered by the user. The default value is
0.8% but can be set to the required value in %.
Summary of Contents for PRS-330 series
Page 2: ...TEL 516 334 5959 800 899 8438 FAX 516 334 5988 www ietlabs com IET LABS INC...
Page 12: ...PRS 330 Series v Table of Contents Page left intensionally blank...
Page 18: ...SPECIFICATIONS PRS 330 Series 6 Page left intensionally blank...
Page 34: ...OPERATION PRS 330 Series 22 Page left intensionally blank...
Page 36: ...USB INTERFACE PRS 330 Series 24 Page left intensionally blank...
Page 44: ...PRS 330 Series ETHERNET INTERFACE OPTION 32...
Page 48: ...PROGRAMMING PRS 330 Series 36 Page left intensionally blank...
Page 67: ...PRS 330 Series APPENDIX B 55...