PRS-330 Series
APPENDIX B
51
SOURCE Commands
SOURce:DATA <display value>
Sets output resistance of the PRS-330 if table is set to
None or the resistance value corresponding to the display value for the table selected.
Display value in form of number and decimal point representing desired display value.
example SOURce:DATA 1.000002 would set resistance value to 1.000002
Ω
if Table is set to
None.
SOURce:DATA?
Returns display value currently shown in the PRS display
SOURce:RESistance <display value> Sets output resistance of the PRS-330 if table is set to
None or the resistance value corresponding to the display value for the table selected.
Display value is in the form of number and decimal point representing desired display value.
This is the same as SOURce:DATA command.
example SOURce:RES 0 would set display value to 0 and if Table is set to PT-100 or PT-1000 this
would represent the resistance value depending upon how the table is con
fi
gured. If in Celsius and PT-
100 then resistance of 100
Ω
would be set at output of PRS-330.