2-5
2
that corresponds to the model that you are configuring and press
Return
e.g.
4863 <return>
for the Model 4803
2363 <return>
for the Model 2303
6.
Cycle Power
The program may ask that you turn the unit's power off and on.
Press Return when the unit has finished its self test.
7.
GPIB Address
The program will ask for the unit's current GPIB address. For a
primary address, enter the address as a one or two digit primary
address. For new units, use the factory default address of 4. For a
secondary address, use 4 digits in the ppss format (2 digit primary,
2 digit secondary address, i.e. 0401).
Note - If you do not know your unit's GPIB address, turn the unit
off and back on. The unit will display its primary GPIB address at
the end of its self test routine by blinking its LEDs. A double blink
indicates the unit is in the dual secondary address mode. The LED
bit weights are:
RDY TALK LSTN SRQ ERR
16
8
4
2
1
8.
Configuration Choices
The configuration program steps through each configurable function,
displays the current setting and gives you a multiple choice or asks
for a numeric entry. Refer to the SCPI commands in Tables 3-2 and
3-3 for an explanation of the choices and entry values. All choices
are made by entering the corresponding choice or value and
pressing the Return key. To skip a parameter, just press the Return
key.
Note that the configuration program is only able to set the digital
interface for string type commands. Settings for direct port I/O
commands or for the Status Reporting Structure will have to be