5-4
5
TABLE 5-1 4863/2363 - SELF TEST ERROR CODES
CONTINUED
Blinking
LED
Error
Possible Fault
SRQ
GPIB
Defective GPIB controller chip. Replace
GPIB Controller, U4 in 4863.
ERR
Address
Address value should be between 0 and 30.
Setting
Check and or correct address setting.
SRQ
GPIB
Wrong firmware or GPIB chip for hardware
+
configuration. Check hardware configuration
ERR
and change firmware or replace GPIB
Controller chip, U4 with NI 7210.
LSTN
CPU
Wrong CPU type. Should be Zilog
+
Z8S18020VSC (SL1919 Enhanced Version)
ERR
TALK
CPU
Wrong CPU type. Should be Zilog
+
Z8S18020VSC
LSTN
+
ERR
RDY
PLD U8
Wrong decoder PLD, U8, or wrong version
+
PLD
LSTN
+
ERR