3-29
3
Notes -
(1)
Setting changes are immediate after command execution.
Provide 15 msec. delay for internal reconfiguration.
(2)
Setting other than the default value are not saved permanently.
INITiate
Starts Trigger branch
[:IMMediate]
Enables a single trigger operation
:CONTinuous
OFF
Enables ongoing external triggers. Values
= 0|1 or OFF|ON.
ABORt
Disables trigger function
CALibrate
Starts calibrate branch
:IDN <string>
ICS Electronics
Sets user IDN message. String is up to 72
characters and consists of four fields (manu-
facturer, model code, serial number and
firmware revision) separated by commas.
e.g. ICS Electronics,4863, S/N 012345,
Rev 0.6 (05-19-06).
:DATe <date>
00/00/0000
Saves IDN message and date. The save
operation lights all the LEDs. Date is in
mm/dd/yy format.
:DATe?
Queries the calibration date. The response
is 00/00/0000 for factory default settings.
DEFault
Sets Flash to factory settings.
:LOCK
0
Disables configuration commands when On.
Values = 0|1 or OFF|ON. Tables 1-4 and 1-
5 list the locked commands.
TABLE 3-3 SCPI COMMANDS AND QUERIES
(CONTINUED)
Keyword
Default
Description
Value