ContaminationSensor CS 1000
Status messages
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 86 / 108
BeWa CS1000 3247149p 240 en-us 2011-02-11.doc
2011-02-11
Status messages
Status LED / Display
LED Blink code /
Display /
Analog output /
Switch out
Status
To do
Error
no.
Gree
n
Conductive
CS o.k.
---
-
Red
<)<(</
2CLEAN
Current value mA / V*
Conductive
The sensor is below
its measurement
range ISO 9/8/7
---
-
Error
LED Blink code /
Display /
Analog output /
Switch out
Status
To do
Error
no.
Red
2 LOW
4.4 mA / 2.2 V*
open
Flow is too low.
Check that the flow is
between 30 … 300
ml/min.
Increase the inlet
pressure or reduce
the outlet pressure.
1
Red
CHECK
19.9 mA / 9.95 V*
open
It is not possible to
determine the flow
rate.
The sensor status is
undefined.
Check that the flow is
between 30 … 300
ml/min.
If the fluid cleanliness
is below the
measurement limit
(ISO 9/8/7), it may
take several
measurement cycles
until measured
values are displayed.
3