ContaminationSensor CS 1000
Analog output
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 84 / 108
BeWa CS1000 3247149p 240 en-us 2011-02-11.doc
2011-02-11
Fluid temperature TEMP
The current range 4.8 … 19.2 mA or voltage range 2.4 … 9.6 V is dependent on the
fluid temperature of from -25°C … 100°C (resolution: 1°C) or -13°F … 212°F
(resolution: 1°F)
Current I
Temperature / error
Voltage U
I < 4.00 mA
Cable break
U < 2.00 V
4.0 mA < I < 4.1 mA
Device error, device not ready
2.00 V < U < 2.05 V
4.1 mA < I < 4.3 mA
Not defined
2.05 V < U < 2.15 V
4.3 mA < I < 4.5 mA
Flow error
(flow insufficient)
2.15 V < U < 2.25 V
4.5 mA < I < 4.8 mA
Not defined
2.25 V < U < 2.40 V
I = 4.8 mA
-25 °C / -13 °F
U = 2.40 V
…
…
…
I = 7.68 mA
0 °C / 32 °F
U = 3.84 V
I = 8.26 mA
+5 °C / 41 °F
U = 4.13 V
I = 8.83 mA
+10 °C / 50 °F
U = 4.42 V
I = 9.41 mA
+15 °C / 59 °F
U = 4.70 V
I = 9.98 mA
+20 °C / 68 °F
U = 4.99 V
I = 10.56 mA
+25 °C / 77 °F
U = 5.28 V
I = 11.14 mA
+30 °C / 86 °F
U = 5.57 V
I = 11.71 mA
+35 °C / 95 °F
U = 5.86 V
I = 12.29 mA
+40 °C / 104 °F
U = 6.14 V
I = 12.86 mA
+45 °C / 113 °F
U = 6.43 V
I = 13.44 mA
+50 °C / 122 °F
U = 6.72 V
I = 14.02 mA
+55 °C / 131 °F
U = 7.01 V
I = 14.59 mA
+60 °C / 140 °F
U = 7.30 V
I = 15.17 mA
+65 °C / 149 °F
U = 7.58 V
I = 15.74 mA
+70 °C / 158 °F
U = 7.87 V
I = 16.32 mA
+75 °C / 167 °F
U = 8.16 V
I = 16.90 mA
+80 °C / 176 °F
U = 8.45 V
I = 17.47 mA
+85 °C / 185 °F
U = 8.74 V
I = 18.05 mA
+90 °C / 194 °F
U = 9.02 V
I = 18.62 mA
+95 °C / 203 °F
U = 9.31 V
I = 19.20 mA
+100 °C / 212 °F
U = 9.60 V
19.2 mA < I < 19.8 mA
Not defined
9.60 V < U < 9.90 V
19.8 mA < I < 20 mA
No measured value
9.90 V < U < 10 V