ContaminationSensor CS 1000
Analog output
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 67 / 108
BeWa CS1000 3247149p 240 en-us 2011-02-11.doc
2011-02-11
HDA.ISO – Analog signal ISO to HDA 5500
The HDA.ISO signal consists of 4 measured values (START / ISO 6 / ISO 6 / ISO 14
/ ISO 21 / Status) which are output sequentially. Synchronization with the
downstream control unit is a prerequisite.
The signal output is as follows:
Time
Measured variable
Signal duration
in s
Current (I) / Voltage (U)
Start signal 0
--
2
20 mA / 10 V
Pause
2
4 mA / 2 V
Signal 1
ISO 4
2
Current/Voltage for signal
Pause
2
4 mA / 2 V
Signal 2
ISO 6
2
Current/Voltage for signal
Pause
2
4 mA / 2 V
Signal 3
ISO 14
2
Current/Voltage for signal
Pause
2
4 mA / 2 V
Signal 4
ISO 21
2
Current/Voltage for signal
Pause
2
4 mA / 2 V
Signal 5
Status
2
Current/Voltage for signal
Pause
30
4 mA / 2 V