ContaminationSensor CS 1000
Operating the CS1000 using the keyboard (only CS1x2x)
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 36 / 108
BeWa CS1000 3247149p 240 en-us 2011-02-11.doc
2011-02-11
Deactivate display FREEZE
If FREEZE is set to MANUAL in the PowerUp menu:
Press both buttons simultaneously to return to the current display.
The display switches to the preset display.
All values present in the FREEZE memory are deleted.
If FREEZE is set to TIMEOUT in the PowerUp menu:
You are returned automatically to the current display after 10 times the value for
MTIME, or manually by pressing both arrow keys simultaneously.
The factory setting of MTIME is 60 seconds * 10 = 600 seconds = 10 minutes.
Modes and menus
The sensor has the following two operating levels / menus.
Menus
Mode
Description
Page
PowerUp Menu
PowerUp Mode
You carry out the basic
settings in this menu.
Measuring Menu Measurement mode
This menu starts automatically
after powering up.
PowerUp Menu
You can carry out the basic settings for operation of the CS in the PowerUp menu.
Selection
To do
Start the PowerUp menu
Press any key while the supply voltage to the
sensor is switched on / generated.
Exit the PowerUp menu without
saving
Scroll through to
CANCEL
and press the
o.k.
-key.
If a key is not pressed within 30 seconds, the
system jumps back automatically.
Exit the PowerUp menu after
saving.
Scroll through to
SAVE
and press the
o.k.
-
key.
PowerUp Menu:
Designation
MODE
Select measurement mode
mTIME
Set measuring duration
pPRTCT
Set pump protection time
ADRESS
Set bus address