ContaminationSensor CS 1000
Analog output
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 71 / 108
BeWa CS1000 3247149p 240 en-us 2011-02-11.doc
2011-02-11
Current I
ISO code / error
Voltage U
I = 15.62 mA
ISO 19
U = 7.81 V
I = 16.19 mA
ISO 20
U = 8.10 V
I = 16.76 mA
ISO 21
U = 8.38 V
I = 17.33 mA
ISO 22
U = 8.67 V
I = 17.90 mA
ISO 23
U = 8.95 V
I = 18.47 mA
ISO 24
U = 9.24 V
I = 19.20 mA
ISO 24.28
U = 9.60 V
19.2 mA < I < 19.8 mA
Not defined
9.60 V < U < 9.90 V
19.8 mA < I < 20 mA
No measured value
9.90 V < U < 10 V
The current (I) or voltage (U) can be calculated for a given ISO contamination
class as follows:
I = 4.8 mA + ISO code x (19.2 mA - 4.8 mA) / 24.28
U = 2.4 V + ISO Code x (9.6 V - 2.4 V) / 24.28
The ISO contamination class can be calculated for a given current I or voltage U
as follows:
ISO code = (I - 4.8 mA) x (24.28 / 14.4 mA)
ISO code = (U - 2.4 V) x (24.28 / 7.2 V)
ISO 2 / ISO 5 / ISO 15
The ISO x setting enables the value of a class to be continuously output via the
analog output.