70 Auto Ref/Keratometer HRK-7000/7000A --------------------------------------------------------
4.
Adjustment of measurement position and focus
-
Ask the examinee to look at the internal fixed target.
-
Adjust the position so that the pupil shall be in between two(2)
perpendicular bars by moving the operation lever.
-
Adjust the focus so that the corneal corner can be seen apparently.
NOTE
As adjusting the focus on the iris, it is impossible to measure the radius of pupil exactly.
5.
Measurement
-
As pushing the measurement button, the window shall be stopped.
-
button and button in the center is to adjust the movement of left bar,
and button and button in right side is to adjust the movement of right bar.
-
Move the relevant bar to left/right sides by pushing button or button.
-
The measured value shall be indicated on the screen of monitor.
-
Store the measured value by pushing the measurement button.
-
The measured value is to be inputted beside “1” of right bottom of the screen. It
is to be inputted beside “AVG” of left bottom of the screen as well.
-
As pushing MODE button, the stopped window is cancelled, and “2” of left
bottom of the screen is to be selected as the bar. Every time pushing MODE
button, “1” or ”2” is to be selected alternatively. If there was an error in “1” which
is the previous measurement, you can select “1” again.
Summary of Contents for HRK-7000
Page 1: ... Auto Ref Keratometer HRK 7000 7000A 1 User s Manual Auto Ref Keratometer HRK 7000 7000A ...
Page 43: ... Auto Ref Keratometer HRK 7000 7000A 43 Figure 22 Example of Print ...
Page 47: ... Auto Ref Keratometer HRK 7000 7000A 47 Figure 24 Example of Print ...
Page 53: ... Auto Ref Keratometer HRK 7000 7000A 53 Figure 26 Screen indicating KER P Mode ...
Page 74: ...74 Auto Ref Keratometer HRK 7000 7000A Figure 41 Display of VIRTUAL COMPARISON Mode ...