![Huvitz HRK-7000 User Manual Download Page 41](http://html1.mh-extra.com/html/huvitz/hrk-7000/hrk-7000_user-manual_2190255041.webp)
--------------------------------------------------------Auto Ref/Keratometer HRK-7000/7000A 41
8.2.
Keratometry (KER Mode)
It is the mode to measure the corneal curvature solely.
Do not measure the base curve of hard contact lens in this mode. Please refer to
CLBC mode in section 8.5 regarding the base curve of hard contact lens.
1.
Check whether or not the screen of monitor is in measurement mode.
2.
KER Mode Selection
-
Push MODE button until “KER” is to be indicated on the upper right side of the
screen.
3.
Perform the same 2, 3 procedure of consecutive measurement of refractive power.
8.2.1.
Manual Measurement Mode
①
Perform the adjustment of array and focus as like in the procedure of section 8.1.1.
②
Measurement
-
Push the measurement button.
-
The measurement continues to be performed as you keep pushing the
measurement button.
-
As the measurement is completed, the measured result is to be indicated on the
screen of monitor. In case of the consecutive measurement, the result of
previous measurement is to be indicated.