86 Auto Ref/Keratometer HRK-7000/7000A --------------------------------------------------------
10.
Self diagnosis & Maintenance
10.1.
Before calling for serviceman
In case that abnormality happens or the machine operates abnormally, a warning sign
is to be indicated. In this case, perform the settlements below.
If the machine does not return to the normal condition in spite of the measures below,
contact to the agent where you bought the machine after switching the power off.
①
As the power switch is on
Message
Cause
Method of settlement
Motor Error
Internal abnormality for the
equipment
Re-input the power in 10
seconds after switching it
off. In case that the
message is indicated
again, contact our sales
representative.
EEPROM Error
EEPROM Data Error
System Error
Clock Error
INVALID SETUP DATA —
REF
Abnormality in the internal
data for Refractometry
Please contact our sales
representative.
INVALID SETUP DATA -
KER
Abnormality in the internal
data for Keratometry
Please contact the selling
agent.