Service/maintenance
7.1
Electrical fuse (if available)
To find out whether your temperature control unit is equipped with fuses, refer to the “Wiring
diagram”.
→
The overcurrent circuit breakers for all pole breaking (L and N) are located at the back of the tem-
perature control unit. In case of a fault (no function and no display on the temperature control unit)
please first check if the overcurrent circuit breaker has tripped. If the overcurrent circuit breakers
immediately trip again after the reset: Unplug the power cord and immediately contact Customer
Support.
→
7.2
Display of errors - only for MPC controllers
The device displays an alarm or warning message in the event of a fault.
Display
Cause
Effect, measure
Flashing display of the
temperature value
Warning: Over or under temperature.
Control continues.
Setpoint limits: ± 2 K
F1 - flashing
Error Sensor1
Break or short circuit
Control is inactive.
(Pump off, compressor off, heating off)
Check the sensor.
E1 - flashing
Input E1 reports an error.
a) No enable signal, level alarm
b) Valid for temperature control units
with heater: The internal temperature is
above the set value of the overtempera-
ture protection. The overtemperature
protection was triggered.
a) Control is inactive. (Pump off, compres-
sor off, heating off)
Check level.
Restart only possible when
level OK.
b) The value of the overtemperature
protection must be above the internal
temperature and/or the setpoint. Do not
set the setpoint above the temperature set
for the overtemperature protection.
E2 - flashing
Input E1 reports an error.
a) Pump runs and no flow or pump is
running and no water pressure.
b) Valid for temperature control units
with heater: The internal temperature is
above the set value of the overtempera-
ture protection. The overtemperature
protection was triggered.
a) Control is inactive. (Pump off, compres-
sor off, heating off)
Restart only possible by interrupting the
power supply.
b) The value of the overtemperature
protection must be above the internal
temperature and/or the setpoint. Do not
set the setpoint above the temperature set
for the overtemperature protection.
E3 - flashing
Although the control is off, the input E1
signals a flow
Control is inactive.
(Pump off, compressor off, heating off)
The error is corrected automatically when
input E1 is open in standby again.
EP - flashing
Loss of data in the parameter memory
Control is inactive.
(Pump off, compressor off, heating off)
Overview of
messages
Summary of Contents for MPC TC 100
Page 1: ...MPC Immersion Cooler...
Page 2: ......
Page 3: ...OPERATION MANUAL MPC Immersion Cooler...
Page 4: ......
Page 6: ...OPERATION MANUAL Immersion Cooler V2 3 0en 03 08 21 1 30 6 MPC controller...
Page 10: ...OPERATION MANUAL Immersion Cooler V2 3 0en 03 08 21 1 30 10...
Page 46: ...Annex OPERATION MANUAL Chapter 9 Immersion Cooler V2 3 0en 03 08 21 1 30 46 9 Annex...
Page 47: ......