Commissioning
Chapter 2
OPERATION MANUAL
V2.3.0en/03.08.21//1.30
Immersion Cooler
23
▪
Use only indoors. The illuminance must be at least 300 lx.
▪
Installation altitude up to 2,000 meters above sea level.
▪
Maintain wall and ceiling clearance for adequate air exchange (dissipation of waste heat, supply
of fresh air for the temperature control unit and work area). Ensure adequate floor clearance for
air-cooled temperature control units. Do not operate this temperature control unit from within
the box or with an inadequately dimensioned bath. This inhibits the air exchange.
▪
Ambient temperature values are provided on the technical data sheet; to ensure trouble-free
operation, compliance with the ambient conditions is mandatory.
▪
Relative humidity max 80% to 32 °C and 40 °C decreasing linearly to 50%.
▪
Short distance to supply connections.
▪
The temperature control unit must not be installed so as to hinder or even prevent access to the
disconnecting device (to the power supply).
▪
For the magnitude of the mains voltage fluctuations, refer to the datasheet.
→
From page 46,
section
▪
Transient surges, as would normally occur in the power supply system.
▪
Installation Class 3
▪
Applicable degree of soiling: 2.
▪
Surge category II.
Please also note:
→
»Exemplary illustrations of the cooling variants«
Side
Distance in cm
[A2]
Top
free standing
[B]
Left
min. 20
[C]
Right
min. 20
[D]
Front
min. 20
[E]
Rear
min. 20
Side
Distance in cm (for operation in a bath)
[A2]
Top
free standing
[B]
Left
min. 20
[C]
Right
min. 20
[D]
Front
min. 20
[E]
Rear
min. 20
Wall clearances
Summary of Contents for MPC TC 100
Page 1: ...MPC Immersion Cooler...
Page 2: ......
Page 3: ...OPERATION MANUAL MPC Immersion Cooler...
Page 4: ......
Page 6: ...OPERATION MANUAL Immersion Cooler V2 3 0en 03 08 21 1 30 6 MPC controller...
Page 10: ...OPERATION MANUAL Immersion Cooler V2 3 0en 03 08 21 1 30 10...
Page 46: ...Annex OPERATION MANUAL Chapter 9 Immersion Cooler V2 3 0en 03 08 21 1 30 46 9 Annex...
Page 47: ......