Commissioning
Chapter 2
OPERATION MANUAL
V2.3.0en/03.08.21//1.30
Immersion Cooler
27
2.8
Preparations for operation
2.8.1
Unscrewing/activating the leveling feet (if any)
The leveling feet are not unscrewed/activated before switching on the temperature control unit
DEATH OR SERIOUS INJURY DUE TO CRUSHING
The parking brakes must be activated at the casters (if any) and/or the leveling feet must be
unscrewed/activated before the temperature control unit is put into operation.
The temperature control unit may move if the parking brakes of the casters (if any) are not
activated and/or the leveling feet are not unscrewed/activated.
Always unscrew/activate the leveling feet before switching on the temperature control unit.
Uneven floors can be compensated by adjusting these leveling feet.
PROCEDURE
Verify that the parking brakes of the casters (if any) have been activated.
Unscrew the leveling feet.
Compensate uneven floors by adjusting these leveling feet, if necessary. Use a spirit level to
horizontally align the temperature control unit.
Tighten the lock screws on the leveling feet after aligning the temperature control unit. This
prevents the leveling feet from changing their height during operation.
2.8.2
Use of the Probe [67]
Observe the wiring diagram.
→
The >Probe< [67] is touched in icy condition
SEVERE FROSTBITES CAUSED BY HOUSING PARTS
The
>Probe<
[67]
must not
be touched in icy condition.
Wear your personnel protective equipment (e.g. temperature-resistant safety gloves, safety
goggles, safety footwear).
The protective and insulating hose is twisted or bent
DAMAGE TO THE TEMPERATURE CONTROL UNIT
Ensure the protective and insulating hose is never bent or twisted.
The protective and insulating hose is immersed into the thermofluid
DAMAGE TO THE TEMPERATURE CONTROL UNIT
Ensure the protective and insulating hose is not immersed into and does not come into contact
with the thermofluid.
The cooling coil is moved / bent when cold.
MATERIAL DAMAGE CAUSED BY INTERNAL FRACTURE
The cooling coil may only be moved / bent when the cooling coil temperature is at room temperature.
The minimum bending radius of the hose is 400 mm and
must never
be smaller.
Valid for temperature control units with MPC controller:
Controlling the temperature requires the Pt100 process control sensor to be connected to the sock-
et
>Pt100 Process<
[49]). Please note the information regarding the interface.
→
Page 36, section
»Interfaces and software update«
Summary of Contents for MPC TC 100
Page 1: ...MPC Immersion Cooler...
Page 2: ......
Page 3: ...OPERATION MANUAL MPC Immersion Cooler...
Page 4: ......
Page 6: ...OPERATION MANUAL Immersion Cooler V2 3 0en 03 08 21 1 30 6 MPC controller...
Page 10: ...OPERATION MANUAL Immersion Cooler V2 3 0en 03 08 21 1 30 10...
Page 46: ...Annex OPERATION MANUAL Chapter 9 Immersion Cooler V2 3 0en 03 08 21 1 30 46 9 Annex...
Page 47: ......