1.4.3
Obligations of the operators
Carefully read the operation manual before operating the temperature control unit. Please observe
the safety instructions. When operating the temperature control unit, wear appropriate personal
protective equipment (e.g. safety goggles, protective gloves, non-slip shoes).
1.5
General information
1.5.1
Description of workstation
The workstation is located at the control panel in front of the temperature control unit. The workstation is
determined by the customer's connected peripheries. Accordingly, it must be designed safe by the re-
sponsible body. The workstation design also depends on the applicable requirements of the German
occupational health and safety regulations [BetrSichV] and the risk analysis for the workstation.
1.5.2
Safety devices to DIN 12876
The temperature control unit is operated with a heating element without additional protection.
RISK OF INJURY
The temperature control unit does
not
have an overtemperature protection and must be
addi-
tionally protected
when used with heating elements.
Note the temperature control unit’s maximum operating temperature. For the values, refer to
the datasheet.
→
The rating of your temperature control unit is stated on the data sheet in the appendix.
Classification
Temperature
control medium
Technical requirements
Identification
d)
I
Non-combustible
a)
Overheat protection
c)
NFL
II
Combustible
b)
Adjustable overheat protection
FL
III
Combustible
b)
Adjustable overtemperature protection
and additional low-level protection
FL
a)
Usually water; other fluids only if non-combustible even within the temperature range of an individual fault.
b)
The temperature control media must have a fire point of ≥ 65 °C.
c)
The overheat protection can, for instance, be realized using a suitable fill level sensor or a suitable temperature limiter.
d)
Optional at the choice of the manufacturer.
▪
Temperature control units with heating correspond to class number III/FL. These temperature
control units are characterized by an “H” in the device name.
▪
Temperature control units without heating correspond to class number I/NFL.
Rating of laboratory
thermostats and
laboratory baths
Overview of the tem-
perature thresholds.
The setpoint can only
be changed for tem-
perature control units
with MPC controller
Summary of Contents for MPC TC 100
Page 1: ...MPC Immersion Cooler...
Page 2: ......
Page 3: ...OPERATION MANUAL MPC Immersion Cooler...
Page 4: ......
Page 6: ...OPERATION MANUAL Immersion Cooler V2 3 0en 03 08 21 1 30 6 MPC controller...
Page 10: ...OPERATION MANUAL Immersion Cooler V2 3 0en 03 08 21 1 30 10...
Page 46: ...Annex OPERATION MANUAL Chapter 9 Immersion Cooler V2 3 0en 03 08 21 1 30 46 9 Annex...
Page 47: ......