Setup mode
OPERATION MANUAL
Chapter 4
Immersion Cooler
V2.3.0en/03.08.21//1.30
34
4
Setup mode
4.1
Setup mode
Moving the temperature control unit during operation
SERIOUS BURNS/FREEZING OF THE HOUSING PARTS/ESCAPING THERMAL FLUID
Do not move temperature control units that are in operation.
The cooling coil is moved / bent when cold.
MATERIAL DAMAGE CAUSED BY INTERNAL FRACTURE
The cooling coil may only be moved / bent when the cooling coil temperature is at room tem-
perature.
4.1.1
Turning on the temperature control unit - without controller
PROCEDURE
Switch on the temperature control unit using the
>Mains isolator<
[36]/
>Mains switch<
[37].
Temperature control starts
immediately
for single-stage cooling machines (up to TC50), in a two-
stage cooling machine (from TC100) temperature control begins shortly after turning on the
temperature control unit. The temperature is lowered until the heat input to the probe corre-
sponds to the cooling capacity of the temperature control unit.
4.1.2
Turn temperature control unit off - without controller
PROCEDURE
Switch off the temperature control unit using the
>Mains isolator<
[36]/
>Mains switch<
[37].
The temperature control stops
immediately
.
4.1.3
Turning off the temperature control unit / with controller
PROCEDURE
Switch on the temperature control unit using the
>Mains isolator<
[36]/
>Mains switch<
[37].
Thermoregulation is
off
. The cooling capacity of the temperature control unit is available only 6
minutes after switch-on.
4.1.4
Turning off the temperature control unit - with controller
The power supply is interrupted before the temperature control process was terminated proper-
ly
DAMAGE TO THE TEMPERATURE CONTROL UNIT
End the temperature control process before the power supply is interrupted (by switching off
or disconnecting from the voltage).
Do not switch off the temperature control unit while the temperature control process is running.
Only switch off the temperature control unit after the temperature control process is finished.
→
Page 35, section
»Ending the temperature control process - with controller«
PROCEDURE
Switch off the temperature control unit using the
>Mains isolator<
[36]/
>Mains switch<
[37].
Switch off the temperature control unit only when
no
temperature control process is activated!
→
»Ending the temperature control process - with controller«
Summary of Contents for MPC TC 100
Page 1: ...MPC Immersion Cooler...
Page 2: ......
Page 3: ...OPERATION MANUAL MPC Immersion Cooler...
Page 4: ......
Page 6: ...OPERATION MANUAL Immersion Cooler V2 3 0en 03 08 21 1 30 6 MPC controller...
Page 10: ...OPERATION MANUAL Immersion Cooler V2 3 0en 03 08 21 1 30 10...
Page 46: ...Annex OPERATION MANUAL Chapter 9 Immersion Cooler V2 3 0en 03 08 21 1 30 46 9 Annex...
Page 47: ......