Function description
Chapter 3
OPERATION MANUAL
V2.3.0en/03.08.21//1.30
Immersion Cooler
31
3.3
To be noted when planning the test
Observe the intended operation.
→
Page 13, section
The focus is on your application. Bear in mind that the system performance depends on the temperature.
▪
Make sure that the electrical connection is adequately dimensioned.
▪
The installation location of the temperature control unit should be selected so as to ensure ade-
quate fresh air, even with water-cooled chillers.
▪
The thermal fluid used must be chosen to not only allow the minimum and maximum working
temperature but it must also be suitable with regard to the fire point, boiling point and viscosity.
In addition, the thermal fluid must be compatible with all the materials in your system.
▪
Avoid bending the cooling coil and cooling water hoses (if required). Use suitable angle pieces and
lay the hose connections with a large radius. Take the minimum bending radius from the data
sheet of the coolant hoses used.
▪
Avoid bending/moving the cooling coil in a cold state.
▪
Check the hoses at regular intervals for any material fatigue (e.g. cracks, leaks).
▪
Water as well as water and anti-freeze mixes must not be used as thermal fluids!
▪
Basically, you should only use the thermal fluids recommended by the manufacturer and only
within the usable temperature and pressure range.
For water-cooled temperature control units, please take the cooling water temperature necessary
for perfect operation and the required differential pressure from the data sheet.
→
From page 46,
3.4
Only valid for temperature control units with MPC controller:
3.4.1
Display and control instruments
3.4.1.1
Display
The value of the internal temperature is displayed. For example, this is the bath temperature for
temperature control units with a bath, or the flow temperature of chillers. Different key combina-
tions also display the setpoint, a menu item or settings.
3.4.1.2
LED display status
These LEDs indicate the current operating condition.
3.4.1.3
Arrow keys
These keys are used to ( increase or decrease ) the setpoint or select or change a menu item. The
>Arrow keys<
[C] are also required to call the menu up.
3.4.1.4
SET key
The
>SET key<
[D] is used to switch to the setpoint temperature. It is used to display and modify the set-
point temperature. The
>SET key<
[D] is also used to display the values of the various menu items.
The control panel:
Displays and keys
Summary of Contents for MPC TC 100
Page 1: ...MPC Immersion Cooler...
Page 2: ......
Page 3: ...OPERATION MANUAL MPC Immersion Cooler...
Page 4: ......
Page 6: ...OPERATION MANUAL Immersion Cooler V2 3 0en 03 08 21 1 30 6 MPC controller...
Page 10: ...OPERATION MANUAL Immersion Cooler V2 3 0en 03 08 21 1 30 10...
Page 46: ...Annex OPERATION MANUAL Chapter 9 Immersion Cooler V2 3 0en 03 08 21 1 30 46 9 Annex...
Page 47: ......