Event
code
Message and description
Details and solution
4092
Installed robotic does not support auto
calibration.
This warning event is generated when
newer library firmware attempts to auto
calibrate some robotic assembly types. If
there are no related move events, this
warning can be ignored.
1.
Check for additional events that occurred
at the same time. If there are no
additional events, this warning can be
ignored.
2.
If there are associated move events, run
the slot to slot test.
3.
Verify that the robotic assembly is level
within the module. If the module was
recently moved or the robotic assembly
replaced, the assembly could be out of
alignment.
4.
If the library continues having move
failures, contact service.
4093
Could not obtain an IP address from a
DHCP server.
1.
Check the network configuration settings
from the
Status > Network
screen.
2.
Verify that the DHCP server is reachable
from the library.
3.
Trigger an automatic reconfiguration of
the network interface by changing the
network configuration from the
Configuration > Network
screen or
unplugging the network cable and then
plugging it in after a few seconds.
4094
Drive interface I/O error.
Reboot the library to reinitialize the
hardware and device drivers.
4095
Library test failed. Not enough valid
cartridges available for testing.
1.
Review the cartridge requirements for
the test and then ensure that sufficient
cartridges are available in the required
locations to run the test.
2.
Rerun the test.
Table Continued