MKP
MKP?
:CALCulate:MARKer[1]|3|3|4:X:POSition <integer>
:CALCulate:MARKer[1]|3|3|4:X:POSition?
Places the active marker at the given
x coordinate.
The HP 8590-series analyzer outputs data in
the format: 200. The HP ESA series analyzer
outputs data in the format:
+2.00002.
MKPAUSE
Pauses the sweep at the active marker for the
duration of the delay period.
MKPK
MKPK HI
MKPK NL
MKPK NH
MKPK NR
:CALCulate:MARKer[1]|2|3|4:MAXimum
:CALCulate:MARKer[1]|2|3|4:MAXimum:LEFT
:CALCulate:MARKer[1]|2|3|4:MAXimum:NEXT
:CALCulate:MARKer[1]|2|3|4:MAXimum:RIGHt
Positions the active marker on a signal peak.
Places the selected marker on the highest point
on the trace that is assigned to that particular
marker number.
Places the selected marker on the next highest
signal peak to the left of the current marked
peak.
Places the selected marker on the next highest
signal peak from the current marked peak.
Places the selected marker on the next highest
signal peak to the right of the current marked
peak.
HP 8590-Series
Analyzers
Command(s)
Corresponding SCPI Command(s)
Description/Comments
Summary of Contents for E4401B
Page 6: ...vi ...
Page 14: ...xiv Contents ...
Page 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Page 65: ...3 1 3 Programming Examples ...
Page 151: ...4 1 4 Programming Command Cross References ...
Page 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Page 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Page 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Page 368: ...7 1 7 Error Messages ...
Page 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...